Purificatori di Gas

Products

Sostituzioni e kit filtro gas Restek Super Clean

  • High-purity output ensures 99.9999% pure gas (at max. flow of 2 L/min).
  • Designed for easy, hassle-free cartridge changes.
  • Glass inside to prevent diffusion; polycarbonate housing outside for safety.

Visualizza le informazioni del prodotto

Trappola per ossigeno a capacità elevata

  • Removes up to 135 ml of oxygen or 2 g of water.
  • Long life—typically purifies more than 480 cu ft of inert gas.
  • Reduces oxygen to 15 ppb.
  • Maximum operating pressure: 250 psi (1,724 kPa).
  • Flow: 3 L/min @ 32 psi (221 kPa).

Visualizza le informazioni del prodotto

Trappola idrocarburi a livello capillare

  • Packed with an extremely high surface area, baked coconut shell–based activated carbon.
  • Purged with ultra-high-purity helium; ready to use with any carrier gas including hydrogen.
  • Reduces organics to 0.1 ppm (assuming 100 ppm input).
  • Maximum operating pressure: 250 psi (1,724 kPa).

Visualizza le informazioni del prodotto

Indicatore trappola per umidità

  • Reduces water to less than 10 ppb. Indicator changes from yellowish green to blue at 5% relative humidity (suitable for general use).
  • Pre-purged for fast stabilization.
  • Reduces noise from high-sensitivity detectors.
  • Heavy-walled glass body prevents oxygen and water infusion.
  • 10 μm frit prevents microparticulate damage to needle valves and flow controllers.
  • Maximum operating pressure: 100 psi (689 kPa).

Visualizza le informazioni del prodotto

Indicatore trappola per ossigeno

  • Removes oxygen to low ppb levels from He, Ar, N2, or methane.
  • Indicator changes from green to gray when adsorption capacity is depleted.
  • Can be used with nonoxidizing gases such as He, Ar, N2, H2, or CH4.
  • Heavy-walled inner glass tube is protected by a sealed outer plastic tube to prevent oxygen and water infusion.
  • Prepurged for fast stabilization.
  • 100 psi (689 kPa) maximum operating pressure.
  • Reduces oxygen to 100 ppb.
  • 10 µm frits at inlet and outlet.
  • Optimal flow rate: < 150 mL/min.

Visualizza le informazioni del prodotto

Trappole chimiche di sostituzione e parti

per GC Agilent

  • Easy to install.
  • Attach to same fittings as original manufacturer's equipment.

Visualizza le informazioni del prodotto

Kit connettori e trappole per il gas a incastro e in linea Super Clean

  • High-purity output ensures 99.9999% pure gas.
  • Click-On fittings for easy, leak-tight cartridge changes; brass or stainless steel, 1/4" or 1/8".
  • Triple gas trap is ideal for purifying carrier gas—it contains oxygen, moisture, and hydrocarbon scrubbers in one cartridge.
  • Fuel gas trap is ideal for purifying flame ionization detector (FID) fuel gases, removing both moisture and hydrocarbons.
  • Helium-specific triple gas trap is ideal for GC-MS—it contains oxygen, moisture, and hydrocarbon scrubbers in one cartridge.

Visualizza le informazioni del prodotto

Filtri a trappola di sostituzione di grandi dimensioni a incastro e in linea Super Clean

  • Largest-capacity traps available—run longer without replacing.
  • Maximum operating pressure of 11 bar (160 psig).
  • High-purity output ensures 99.9999% pure gas (maximum flow of 8 L/min).
  • Triple gas trap is ideal for purifying carrier gas.
  • Helium-specific triple gas trap is made for GC-MS—ready to use within 15 minutes of installation.
  • Hydrocarbon trap removes hydrocarbon contamination from carrier gas to help eliminate ghost peaks.
Note: Replacement filter only—fittings sold separately.

Visualizza le informazioni del prodotto

Kit per filtro a trappola di grandi dimensioni a incastro e in linea Super Clean

  • Largest-capacity traps available—run longer without replacing.
  • "Quick connect" fittings with spring-loaded check valves allow filter replacement in seconds without tools or contamination.
  • Maximum operating pressure of 11 bar (160 psig).
  • High-purity output ensures 99.9999% pure gas (maximum flow of 8 L/min).
  • Triple gas trap is ideal for purifying carrier gas.
  • Helium-specific triple gas trap is made for GC-MS—ready to use within 15 minutes of installation.
  • Hydrocarbon trap removes hydrocarbon contamination from carrier gas to help eliminate ghost peaks.

Visualizza le informazioni del prodotto

Filtro per ossigeno a capacità ultra-elevata Restek Super Clean

Restek Super Clean ultra-high capacity oxygen filter cartridges feature easy-to-read indicators.

The indicator code is shown on every trap so there is no confusion determining when a trap needs to be replaced.


Visualizza le informazioni del prodotto

Filtro per umidità a capacità ultra-elevata Restek Super Clean

Restek Super Clean ultra-high capacity moisture filter cartridges feature easy-to-read indicators.

The indicator code is shown on every trap so there is no confusion determining when a trap needs to be replaced.


Visualizza le informazioni del prodotto

Trappola idrocarburi riempibile

  • Removes trace impurities from carrier gas.
  • Reduces organics to 0.1 ppm (assuming 100 ppm input).
  • 60 µm frit prevents gas contamination by purifier particles.
  • Good for purge-and-trap systems.
  • Refillable and rechargeable.
  • Maximum operating pressure: 125 psig (862 kPa).
  • Maximum flow: 5 L/min.

Visualizza le informazioni del prodotto

Trappola per umidità a capacità elevata

  • Purged with ultra-high-purity helium; ready to use with any carrier gas including hydrogen.
  • Reduces water to less than 15 ppb.
  • Maximum operating pressure: 250 psi (1,724 kPa).
  • Maximum flow: 1.25 L/min.

Visualizza le informazioni del prodotto

Sistema di gestione gas Restek

  • Removes moisture, hydrocarbons, and oxygen from carrier gas, extending column lifetime.
  • Produces high-purity carrier gas for most applications.
  • Includes one each: moisture, hydrocarbon, and indicating oxygen trap.
  • Replacing traps is safe and easy.
  • Maximum flow: 150 cc/min for optimal performance.

Visualizza le informazioni del prodotto

Carbon Dioxide Trap

  • Removes carbon dioxide from inert gases He, Ar, N2, H2, and clean dry air (CDA) to low ppb levels.
  • Functions by consuming carbon dioxide in a reaction with highly dispersed NaOH on a silicate support (reaction produces water which remains adsorbed in the purifier).
  • In highly critical applications, consider using an additional moisture trap after the CO2 trap as a safety measure.

Visualizza le informazioni del prodotto

Indicating Detector Gas Purifier

  • Removes moisture and hydrocarbons (C5 and heavier) from dry air or inert gases.
  • Indicator signals breakthrough by changing from yellowish green to blue at low relative humidity to warn of leaks in the gas stream.
  • The inner glass tube contains activated carbon, moisture indicator, and molecular sieve.
  • 10 μm frits at both ends of the filter.
  • Heavy-walled inner glass tube is protected by a sealed outer plastic tube to further protect against leaks.

Visualizza le informazioni del prodotto

Indicating Carrier Gas Purifier

  • Removes multiple impurities such as O2, H2O, CO2, CO, halogenated compounds, siloxanes, hydrocarbons (heavier than butane), sulfur-containing molecules, and ammonia from inert gases (He, Ar, N2, and H2).
  • Can be used with other filters to provide endpoint detection.
  • Contains highly sensitive visual indicators for oxygen and water contamination. The oxygen indicator signals breakthrough when the adsorbent changes from green to grey, and the moisture indicator signals breakthrough by changing from blue to brown.
  • Heavy-walled inner glass tube is protected by a sealed outer plastic tube to further protect against leaks.

Visualizza le informazioni del prodotto

Electronic Maintenance Indicators for Gas Filters

  • Audiovisual warning system ensures maintenance is done on time.
  • Progressively displays gas filter status, preventing unplanned downtime for maintenance.
  • Can be custom configured to your replacement/maintenance schedule.

Visualizza le informazioni del prodotto

Dynamic Duo (Leak Detector Elettronico Restek e Flussimetro ProFLOW 6000 Restek)

Proteggete il vostro strumento e migliorate la qualità dei dati grazie a questa accoppiata vincente. Verificare la presenza di perdite ed i flussi vi permetterà di ottimizzare le analisi e prevenire costosi inconvenienti.

Visualizza le informazioni del prodotto

Flussimetro elettronico ProFLOW 6000 Restek

Flussimetri che possono misurare gas infiammabili stanno diventando obbligatori a causa del maggiore uso di idrogeno nella cromatografia. Grazie alla sua "Certificazione Ex", il misuratore Restek ProFLOW 6000 è progettato appositamente per lavorare in modo sicuro in presenza di gas esplosivi e infiammabili.

Restek ProFLOW 6000 è il flussimetro di cui hai bisogno per ogni tipo di misurazione di gas cromatografici grazie alla sua vasta gamma di funzionalità. Il ProFLOW 6000 è un misuratore elettronico in grado di misurare il flusso volumetrico per la maggior parte dei gas. Le misurazioni in tempo reale possono essere effettuate per vari tipi di linee di gas, comprese quelle in cui i gas cambiano continuamente. Questa unità portatile è progettata per un facile utilizzo, e lo stand aggiunge praticità d’uso anche sul bancone.


Visualizza le informazioni del prodotto

Leak Detector Elettronico Restek

Nuovo e migliorato! Non permettete che una piccola perdita vi comporti una costosa riparazione, proteggete la colonna analitica con un Leak Detector Restek.

  • Rileva un’ampia gamma di gas e indica l’entità della perdita con segnali acustici e display LED.
  • Non è più necessario attendere una ricarica completa - può essere utilizzata durante la ricarica o utilizzata fino a 12 ore tra una ricarica e l'altra.
  • Il kit di ricarica include sia l'adattatore di alimentazione CA universale che il cavo di ricarica USB per ricaricare ovunque e in qualsiasi momento.
  • Individuazione di perdite di gas molto piccole in modo rapido e preciso prima che possano causare danni e conseguente arresto delle analisi.
  • L'unità palmare compatta è facile e comoda da usare ovunque sia necessario per controllare eventuali perdite.

Visualizza le informazioni del prodotto

Connettori a incastro e in linea Super Clean

Click-On connectors allow you to change traps quickly, without introducing oxygen into your system.

Visualizza le informazioni del prodotto

Kit e filtro Super Clean specifici per elio

  • Specifically designed for purification of helium in GC-MS systems.
  • Traps are packed and conditioned using helium.
  • Uses standard single-position baseplate.

Visualizza le informazioni del prodotto

Sistema trappola gas Restek Super Clean per LC-MS

U.S. DEA preparazione esente – Non sono richiesti ulteriori permessi al cliente per l’acquisto del prodotto all’interno degli Stati Uniti.

I numeri di registrazione del Canadian Test Kit sono assegnati e disponibili nelle pagine dei singoli prodotti.


Visualizza le informazioni del prodotto

Kit manometro gas

  • Use an in-line pressure gauge to indicate when a thermal gas purifier getter tube should be replaced.
  • Includes 1/8" tee and 0–100 psi (0-689k Pa) gauge.

Visualizza le informazioni del prodotto

Piastre base per filtri Restek

  • End fittings available in brass or stainless steel.
  • Baseplates fit all stand-alone Super Clean gas filters offered.

Visualizza le informazioni del prodotto

Trappola a S per idrocarburi

  • Removes hydrocarbons and other contaminants.
  • Reduces organics to 0.1 ppm (assuming 100 ppm input).
  • Each individually activated to ensure maximum efficiency.
  • Fits in GC oven for easy thermal recharging.
  • Maximum operating pressure: 60 psi (414 kPa).

Visualizza le informazioni del prodotto

Trappola sfiatatoio ECD

  • Reduces the release of hazardous materials from the ECD vent into the lab.
  • Includes connecting lines and mounting kit.

Visualizza le informazioni del prodotto

Moduli depuratore di gas specifico VICI Mat/Sen

  • Replace separate traps for removing impurities (e.g., moisture and hydrocarbons) with one multiple-bed purifier that is designed specifically for purifying helium, hydrogen, nitrogen, or air.
  • Reduce gas impurities from ppm to low ppb levels.
  • Decrease baseline noise and increase GC-MS sensitivity.
  • Pre-purged with the specified gas to shorten downtime.

Visualizza le informazioni del prodotto

Trappola sfiatatoio separata a capacità elevata

  • Reduces the release of hazardous materials from the capillary split vent into the lab.
  • Includes connecting lines and mounting kit.

Visualizza le informazioni del prodotto

Trappola a S setaccio molecolare

  • Traps water vapor; increases column and oxygen trap lifetime.
  • Reduces baseline noise from sensitive detectors such as ECDs and mass spectrometers.
  • Activated and ready to use.
  • Reduces water to less than 1 ppm.
  • Maximum flow: 1 L/min.

Visualizza le informazioni del prodotto

Accessori di Ricarica per il Leak Detector Elettronico Restek

For current leak detector model only (cat.# 28500); not compatible with legacy model (cat.# 22655).

Visualizza le informazioni del prodotto

Adattatore di piccola sonda per rilevatore di perdite elettronico Restek

Compatible with both current (cat.# 28500) and legacy (cat.# 22655) leak detector models.

Visualizza le informazioni del prodotto

Filtri per Gas Compatibili con le basi metalliche Restek Super Clean

  • Purezza del gas in uscita pari a 99.9999% (ad un flusso max di 2 L/min).
  • Studiato per un’installazione facile e senza attrezzi.
  • Il rivestimento interno in vetro previene la diffusione dei gas ed il policarbonato esterno rende il filtro sicuro.
  • Tutti i filtri misurano: 10 5/8" x 1 3/4" (27 x 4.4 cm).

Visualizza le informazioni del prodotto

Astuccio morbido per Rilevatore di perdite o flussimetro ProFLOW 6000

Optional soft-sided storage case is ideal for storing your leak detector or flowmeter in smaller spaces, such as your toolbox.

Compatible with both current (cat.# 28500) and legacy (cat.# 22655) leak detector models.


Visualizza le informazioni del prodotto

Staffa montaggio a parete

Baseplates can be mounted by using screws and the mounting holes on the baseplate, or by using this optional wall mounting bracket.

Visualizza le informazioni del prodotto

Riduttore estremità tubo da 1/8 a 1/4"

To adapt 1/8" Super Clean baseplate fittings to 1/4" tubing, use 1/8" to 1/4" tube end reducers.

Visualizza le informazioni del prodotto

Adattatori di alimentazione per rilevatore di perdite elettronico Restek

For legacy leak detector model only (cat.# 22655); not compatible with current model (cat.# 28500).

Visualizza le informazioni del prodotto