Cannabinoids on Raptor™ ARC-18

Peaks
1.Cannabidivarin (CBDV)
2.Cannabidiolic acid (CBDA)
3.Cannabigerol (CBG)
4.Cannabidiol (CBD)
Peaks
5.Tetrahydrocannabivarin (THCV)
6.Cannabinol (CBN)
7.delta-9-Tetrahydrocannabinol (∆9-THC)
8.Cannabichromene (CBC)
9.delta-9-Tetrahydrocannabinolic acid A (THCA)
image
ColumnRaptor™ ARC-18 (cat.# 9314A65)
Dimensions:150 mm x 4.6 mm ID
Particle Size:2.7 µm
Temp.:50 °C
SampleCannabidiolic acid
Cannabigerol (cat.# 34091)
Cannabidiol (cat.# 34011)
Cannabinol (cat.# 34010)
delta-9-Tetrahydrocannabinol (THC) (cat.# 34067)
Cannabichromene (cat.# 34092)
delta-9-Tetrahydrocannabinolic acid A (THCA) (cat.# 34093)
Diluent:50:50 Methanol:water
Conc.:50 µg/mL
Inj. Vol.:5 µL
Mobile Phase
A:0.1% Formic acid in water
B:0.1% Formic acid in acetonitrile
Time (min)Flow (mL/min)%A%B
0.001.52575
4.001.50100
4.011.52575
7.001.52575
DetectorUV/Vis @ 220 nm
InstrumentHPLC
Download PDF