EPA 8310 PAHs on Pinnacle II PAH by LC-UV

PeakstR (min)Conc.
(µg/mL)
1.Naphthalene2.68100
2.Acenaphthylene3.09100
3.1-Methylnaphthalene3.39100
4.2-Methylnaphthalene3.56100
5.Acenaphthene3.74100
6.Fluorene3.91100
7.Phenanthrene4.40100
8.Anthracene4.93100
9.Fluoranthene5.4910
PeakstR (min)Conc.
(µg/mL)
10.Pyrene5.9110
11.Benzo[a]anthracene7.3910
12.Chrysene7.7710
13.Benzo[b]fluoranthene8.9510
14.Benzo[k]fluoranthene9.525
15.Benzo[a]pyrene9.9910
16.Dibenzo[a,h]anthracene10.7510
17.Benzo[ghi]perylene11.1810
18.Indeno[1,2,3-cd]pyrene11.6010
image
LC_EV0537
ColumnPinnacle II PAH (cat.# 921946E)
Dimensions:150 mm x 3.0 mm ID
Particle Size:4 µm
Pore Size:110 Å
Temp.:30 °C
SampleEPA Method 8310 quality control check (cat.# 31843)
Inj. Vol.:10 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)Flow (mL/min)%A%B
0.001.25050
10.01.20100
12.51.20100
12.511.25050
15.01.25050
DetectorUV/Vis @ 280 nm
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

PAH

,

8310

,

610

,

Polycyclic

,

aromatic

,

hydrocarbons

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.