EPA Method 610 Polycyclic Aromatic Hydrocarbons on Pinnacle DB PAH

Peaks
1.Naphthalene
2.Acenaphthylene
3.1-Methylnaphthalene
4.2-Methylnaphthalene
5.Acenaphthene
6.Fluorene
7.Phenanthrene
8.Anthracene
9.Fluoranthene
Peaks
10.Pyrene
11.Benzo[a]anthracene
12.Chrysene
13.Benzo[b]fluoranthene
14.Benzo[k]fluoranthene
15.Benzo[a]pyrene
16.Dibenzo[a,h]anthracene
17.Benzo[ghi]perylene
18.Indeno[1,2,3-cd]pyrene
image
LC_EV0469
ColumnPinnacle® DB PAH (cat.# 9470252)
Dimensions:50 mm x 2.1 mm ID
Particle Size:1.9 µm
Pore Size:140 Å
Temp.:30 °C
Sample
Diluent:Acetonitrile
Conc.:20 µg/mL each component
Inj. Vol.:2 µL
Mobile Phase
A:Water
B:Acetonitrile
Time (min)%B
0.0050
160
3100
5100
Flow:0.6 mL/min
DetectorUV/Vis @ 254 nm
InstrumentJasco X-LC
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

PAHs

,

610

,

8310

,

UHPLC

,

polycyclic aromatic hydrocarbons

,

EPA Method 610

,

EPA 610

,

polynuclear aromatic hydrocarbons

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.