Creatine and Creatinine in Human Plasma on Raptor HILIC-Si by LC-MS/MS

PeakstR (min)Precursor IonProduct IonProduct Ion
1.Creatinine0.537114.044.386.0
2.Creatine1.232132.143.390.2
image
LC_CF0722
ColumnRaptor HILIC-Si (cat.# 9310A52)
Dimensions:50 mm x 2.1 mm ID
Particle Size:2.7 µm
Pore Size:90 Å
Guard Column:UltraShield UHPLC precolumn filter 0.2 µm (cat.# 25810)
Temp.:40 °C
Sample
Diluent:20:80 Water:acetonitrile
Conc.:Endogenous levels
Inj. Vol.:0.2 µL
Mobile Phase
A:5 mM Ammonium formate in 20:80 water:acetonitrile
Time (min)Flow (mL/min)%A
0.000.5100
1.5Stop
DetectorMS/MS
Ion Mode:ESI+
Mode:MRM
InstrumentUHPLC
NotesSample Preparation:
Endogenous levels of creatinine and creatine in human plasma were determined using a single protein precipitation step followed by LC-MS/MS analysis. A 50 μL aliquot of human plasma (K2EDTA) was mixed with 950 μL acetonitrile. After vortexing and centrifuging at 4,300 rpm for 10 min, 200 μL of the supernatant was transferred to a new vial and mixed with 50 μL of water. Centrifugation was performed again before injection.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

human

,

plasma

,

Raptor

,

HILIC-Si

,

Creatine

,

Creatinine

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.