Antibiotic Cephaloridine on Ultra IBD (LC/MS)

Peaks
1.Cephaloridine
image
LC_0124
ColumnUltra IBD (cat.# 9175552)
Dimensions:50 mm x 2.1 mm ID
Particle Size:5 µm
Pore Size:100 Å
Temp.:Ambient
Sample
Diluent:Mobile phase
Conc.:5 µg/mL
Inj. Vol.:5 µL
Mobile Phase5 mM Ammonium acetate, pH 7.4:acetonitrile (20:80)
Flow:0.2 mL/min
DetectorPE/Sciex API 150 EX
Ion Source:TurboIonSpray®
Ion Mode:ESI+
Ion Spray Voltage:4.7 kV
Orifice Voltage:10.0 V
Ring Voltage:25.0 V
Source Temp.:350 °C
NotesNebulizer: 8L/hour
Curtain gas: 12L/hour
SIM monitoring: 416 ± 3 Daltons
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.