PiONA Calibration Standard on Rtx-DHA-100 by ASTM D6729-14

PeakstR (min)Conc.Wt. %Group
1.3-Methyl-1-butene12.8050.291O5
2.iso-Pentane14.3140.304I5
3.1-Pentene15.891.669O5
4.2-Methyl-1-butene16.7120.42O5
5.n-Pentane17.2051.728P5
6.2-Methyl-1,3-butadiene17.7140.699O5
7.trans-2-Pentene18.2950.542O5
8.cis-2-Pentene19.2940.21O5
9.4-Methyl-1-pentene25.7160.646O6
10.Cyclopentane26.3581.891N5
11.2,3-Dimethylbutane27.1060.447I6
12.2-Methylpentane28.1591.211I6
13.3-Methylpentane30.7511.998I6
14.1-Hexene32.1392.137O6
15.n-Hexane34.441.729P6
16.trans-2-Hexene35.5660.31O6
17.2-Methyl-2-pentene36.0470.654O6
18.cis-2-Hexene37.360.555O6
19.Methylcyclopentane39.1921.184N6
20.2,2-Dimethylpentane39.2780.463I7
21.2,4-Dimethylpentane40.7831.127I7
22.2,2,3-Trimethylbutane41.3120.125I7
23.Benzene44.1411.899A6
24.3,3-Dimethylpentane45.7070.438I7
25.Cyclohexane45.980.954N6
26.2-Methylhexane49.1430.891I7
27.2,3-Dimethylpentane49.1430.887I7
28.3-Methylhexane50.9630.451I7
29.cis-1,3-Dimethylcyclopentane51.7860.06N7
30.trans-1,3-Dimethylcyclopentane52.3461.16N7
31.trans-1,2-Dimethylcyclopentane52.9650.73N7
32.3-Ethylpentane53.1410.168I7
33.1-Heptene53.9411.29O7
34.trans-3-Heptene56.1810.712O7
35.n-Heptane56.5041.723P7
36.cis-3-Heptene56.7220.194O7
37.trans-2-Heptene57.6460.101O7
38.cis-2-Heptene59.2990.649O7
39.Methylcyclohexane59.850.735N7
40.2,2-Dimethylhexane61.270.149I8
41.Ethylcyclopentane62.6421.567N7
42.2,2,3-Trimethylpentane63.3050.698I8
43.2,5-Dimethylhexane63.581.112I8
44.2,4-Dimethylhexane63.9120.276I8
45.ctc-1,2,3-Trimethylcyclopentane66.2820.747N8
46.Toluene67.462.839A7
47.2-Methylheptane70.6550.899I8
48.4-Methylheptane70.9080.286I8
49.3-Methylheptane71.9931.122I8
50.3-Ethylhexane72.1230.058I8
51.trans-1,4-Dimethylcyclohexane72.1230.941N8
52.1-Octene74.6522.155O8
53.1-Ethyl-1-methylcyclopentane74.6520.292N8
54.trans-1,2-Dimethylcyclohexane75.1280.328N8
55.ccc-1,2,3-Trimethylcyclopentane76.0980.159N8
56.n-Octane76.6031.721P8
57.trans-2-Octene77.2150.656O8
58.Isopropylcyclopentane77.4990.261N8
59.cis-2-Octene78.4440.439O8
60.cis-1,2-Dimethylcyclohexane80.0352.055N8
61.1,1,4-Trimethylcyclohexane80.8591.261N9
62.n-Propylcyclopentane82.0270.874N8
63.ctt-1,2,4-Trimethylcyclohexane82.0270.368N9
64.2,5-Dimethylheptane82.2010.733I9
65.3,5-Dimethylheptane82.3980.445I9
66.3,3-Dimethylheptane82.3980.335I9
67.Ethylbenzene83.5011.897A8
68.ccc-1,3,5-Trimethylcyclohexane83.8820.469N9
69.m-Xylene84.720.715A8
70.p-Xylene84.8730.239A8
71.2,3-Dimethylheptane85.0490.22I9
72.3,4-Dimethylheptane85.4110.159I9
73.2-Methyloctane86.2540.245I9
74.ctc-1,2,3-Trimethylcyclohexane86.980.735N9
75.3-Methyloctane87.1211.329I9
76.ctc-1,2,4-Trimethylcyclohexane87.2120.212N9
77.3,3-Diethylpentane87.3890.364I9
78.o-Xylene87.7230.719A8
79.1,1,2-Trimethylcyclohexane87.8290.368N9
80.1-Nonene89.2041.313O9
81.Isobutylcyclopentane89.530.267N9
82.cis-4-Nonene90.0370.829O9
83.trans-3-Nonene90.0370.44O9
84.cis-3-Nonene90.270.575O9
85.n-Nonane90.6411.727P9
86.trans-2-Nonene90.9870.434O9
87.Isopropylbenzene91.9390.48A9
88.2,2-Dimethyloctane92.910.398I10
89.3,3-Dimethyloctane94.9430.153I10
90.n-Propylbenzene95.4070.711A9
91.1-Methyl-3-ethylbenzene96.260.643A9
92.1-Methyl-4-ethylbenzene96.4910.476A9
93.1,3,5-Trimethylbenzene97.1250.097A9
94.2-Methylnonane98.10.3I10
95.1-Methyl-2-ethylbenzene98.20.361A9
96.3-Ethyloctane98.4730.203I10
97.3-Methylnonane98.8071.001I10
98.1,2,4-Trimethylbenzene99.8040.475A9
99.tert-Butylbenzene99.8040.473A10
100.2-Methyl-1-nonene100.0030.648O10
101.Isobutylcyclohexane100.3230.733N10
102.1-Decene100.5540.43O10
103.trans-1-Methyl-2-propylcyclohexane101.460.927N10
104.Isobutylbenzene101.460.713A10
105.n-Decane101.7541.736P10
106.sec-Butylbenzene101.7540.711A10
107.1-Methyl-3-isopropylbenzene102.7860.245A10
108.1-Methyl-4-isopropylbenzene103.1380.949A10
109.1-Methyl-2-isopropylbenzene104.4550.645A10
110.1-Methyl-3-n-propylbenzene105.7380.479A10
111.1-Methyl-4-n-propylbenzene106.1180.477A10
112.n-Butylbenzene106.2120.478A10
113.1,3-Dimethyl-5-ethylbenzene106.3510.094A10
114.1-Methyl-2-n-propylbenzene107.1980.494A10
115.1,4-Dimethyl-2-ethylbenzene108.070.478A10
116.1,2-Dimethyl-4-ethylbenzene108.7470.21A10
117.1,2-Dimethyl-3-ethylbenzene110.4450.495A10
118.1,3-Dimethyl-2-ethylbenzene110.4450.184A10
119.n-Undecane110.7941.723P11
120.1,2,4,5-Tetramethylbenzene111.3940.198A10
121.(2-Methylbutyl)benzene111.560.027A11
122.1-tert-Butyl-2-methylbenzene112.6980.478A11
123.Pentylbenzene114.320.473A11
124.trans-1-Methyl-2-(4-methylpentyl)cyclopentane114.5290.718N12
125.1-t-Butyl-4-ethylbenzene116.0040.478A12
126.n-Dodecane117.6221.727P12
127.1,3,5-Triethylbenzene118.3270.48A12
128.1,2,4-Triethylbenzene119.4840.233A12
129.Hexylbenzene120.6981.186A12
130.n-Tridecane123.1021.723P13
131.5-t-Butyl-1,2,3-trimethylbenzene123.9890.772A13
132.n-Tetradecane127.7191.735P14
133.n-Pentadecane131.7681.734P15
image
GC_PC1330
ColumnRtx-DHA-100, 100 m, 0.25 mm ID, 0.50 µm (cat.# 10148)
SampleDHA PiONA standard (cat.# 30730)
Injection
Inj. Vol.:0.2 µL split (split ratio 200:1)
Liner:Topaz 4.0 mm ID Precision inlet liner w/wool (cat.# 23305)
Inj. Temp.:250 °C
Oven
Oven Temp.:0 °C (hold 15 min) to 50 °C at 1 °C/min to 130 °C at 2 °C/min to 270 °C at 4 °C/min
Carrier GasHe, constant pressure
Linear Velocity:24 cm/sec @ 35 °C
DetectorFID @ 300 °C
Make-up Gas Flow Rate:30 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:370 mL/min
InstrumentAgilent 7890A GC
NotesDetailed hydrocarbon analysis (DHA) is a gas chromatographic technique used to determine the composition of spark ignition fuels. Hydrocarbon streams can be characterized by their individual components or by the group type of hydrocarbon (paraffins = P, isoparaffins = I, olefins = O, naphthenes = N, and aromatics = A). This analysis is a blend of all the group types. To help with identification, individual groups were run under the same conditions and can be found in the following chromatograms:
- GC_PC1322 (isoparaffins)
- GC_PC1323 (aromatic compounds)
- GC_PC1324 (olefins)
- GC_PC1325 (naphthenes)

Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

DHA

,

PiONA

,

PONA

,

gasoline

,

D6729

,

D6730

,

D6733

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.