Florida TRPH Standard Mix on Rxi-1HT

PeakstR (min)
1.n-Octane1.24
2.n-Decane2.71
3.n-Dodecane4.19
4.n-Tetradecane5.49
5.n-Hexadecane6.64
6.n-Octadecane7.68
7.n-Eicosane8.63
8.n-Docosane9.50
PeakstR (min)
9.n-Tetracosane10.30
10.n-Hexacosane11.04
11.n-Octacosane11.73
12.n-Triacontane12.38
13.n-Dotriacontane12.98
14.n-Tetratriacontane13.55
15.n-Hexatriacontane14.09
16.n-Octatriacontane14.60
17.n-Tetracontane15.08
image
GC_PC1272
ColumnRxi-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950)
SampleFlorida TRPH standard (cat.# 31266)
Diluent:Hexane
Conc.:50 ppm (1.25 ng on-column)
Injection
Inj. Vol.:1.0 µL split (split ratio 40:1)
Liner:Premium 4.0 mm ID Precision inlet liner w/ wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:40 °C (hold 1.0 min) to 400 °C at 20 °C/min (hold 10 min)
Carrier GasHe, constant flow
Linear Velocity:55 cm/sec
DetectorFID @ 400 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:50 Hz
InstrumentAgilent/HP6890 GC