Saturated and Unsaturated Hydrocarbons on Rt-Silica BOND PLOT Column

Peaks
1.Methane
2.Ethane
3.Ethylene
4.Acetylene
5.Propane
6.Cyclopropane
7.Propylene
8.Propadiene
Peaks
9.n-Butane
10.1-Butene
11.Methyl acetylene
12.1,3-Butadiene
13.trans-2-Butene
14.Isobutylene
15.cis-2-Butene
16.Isopentane
17.n-Pentane
image
ColumnRt-Silica BOND, 30 m, 0.32 mm ID (cat.# 19785)
SampleCustom DCG gas standard
Diluent:Nitrogen
Conc.:1 mole percent
Injection
Inj. Vol.:15 µL split (split ratio 35:1)
Liner:Premium 2 mm straight inlet liner (cat.# 23313.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:60 °C (hold 2 min) to 175 °C at 2 °C/min
Carrier GasHe, constant flow
Flow Rate:3.3 mL/min
DetectorFID @ 260 °C
Make-up Gas Flow Rate:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:400 mL/min
Data Rate:10 Hz
InstrumentAgilent 7890A GC