Petroleum Oxygenates (ASTM D4815) on TCEP (Micropacked) & Rtx-1

Peaks
1.Methanol
2.Ethanol
3.Isopropanol
4.tert-Butanol
5.n-Propanol
6.Methyl tert-butyl ether
7.sec-Butanol
8.Diisopropyl ether (DIPE)
Peaks
9.Isbutanol
10.Ethyl-tert-butyl ether (ETBE)
11.tert-Amyl alcohol
12.Dimethoxyethane (DME)
13.n-Butanol
14.Benzene
15.tert-Amyl Methyl ether (TAME)
16.total heavy hydrocarbons
image
GC_PC00194
Columns20% TCEP on 80/100 Chromosorb PAW 0.56 m, 0.75 mm ID (cat.# 19040)
and Rtx-1 (connected in series) 30 m, 0.53 mm ID, 3.00 µm (cat.# 10185)
Sampleoxygenates blend
Conc.: 1-10% wt in surrogate gasoline
Injection
Inj. Vol.:0.5 µL split (split ratio 15:1)
Inj. Temp.:200 °C
Oven
Oven Temp.:60 °C
Carrier GasHe, constant flow
Flow Rate:5 mL/min
DetectorFID @ 250 °C
NotesCarrier gas set @ 60°C
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.