GC Multiresidue Pesticide Standard #2-OCP on Rxi-5ms by GC-MS

PeakstR (min)
1.Chloroneb10.10
2.Pentachlorobenzene10.31
3.alpha-BHC12.70
4.Hexachlorobenzene12.90
5.Pentachloroanisole13.02
6.beta-BHC13.36
7.gamma-BHC (Lindane)13.52
8.delta-BHC14.09
9.Endosulfan ether14.55
10.Heptachlor15.25
11.Pentachlorothioanisole15.83
12.Aldrin16.06
13.4,4'-Dichlorobenzophenone16.26
14.Fenson16.45
15.Isodrin16.72
16.Heptachlor epoxide (Isomer B)16.97
17.Chlorbenside17.37
18.trans-Chlordane17.51
19.2,4'-DDE17.62
20.Endosulfan I17.79
PeakstR (min)
21.cis-Chlordane17.85
22.trans-Nonachlor17.96
23.Chlorfenson (Ovex)17.97
24.4,4'-DDE18.31
25.Dieldrin18.37
26.2,4'-DDD18.50
27.Endrin18.85
28.Ethylan (Perthane)18.86
29.Endosulfan II19.04
30.4,4'-DDD19.23
31.2,4'-DDT19.31
32.cis-Nonachlor19.33
33.Endrin aldehyde19.45
34.4,4'-Methoxychlor olefin19.84
35.Endosulfan sulfate19.97
36.4,4'-DDT20.04
37.2,4'-Methoxychlor20.27
38.Endrin ketone20.97
39.Tetradifon21.68
40.Mirex22.16
image
GC_FS0601
ColumnRxi®-5ms, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13423)
SampleGC multiresidue pesticide standard #2-OCP (cat.# 32564)
Diluent:Toluene
Conc.:100 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4.0 mm ID Precision® inlet liner w/wool (cat.# 23305.1)
Inj. Temp.:250 °C
Oven
Oven Temp.:90 °C (hold 1 min) to 330 °C at 8.5 °C/min (hold 5 min)
Carrier GasHe, constant flow
Flow Rate:1.4 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1555-5507
Transfer Line Temp.:290 °C
Analyzer Type:Quadrupole
Source Temp.:325 °C
Electron Energy:70 eV
Solvent Delay Time:5 min
Ionization Mode:EI
InstrumentThermo Scientific TSQ 8000 Triple Quadrupole GC-MS
NotesReconstructed chromatogram from select ions.
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.