Residual Solvents in Cannabis Concentrates on Rxi-624Sil MS by Headspace–Full Evaporation Technique (HS-FET)

PeakstR (min)
1.Isobutane0.903
2.Butane0.989
3.Methanol1.110
4.Pentane1.497
5.Ethanol1.542
6.Acetone1.787
7.Isopropanol1.888
8.n-Hexane2.405
9.Chloroform2.957
10.Benzene3.208
11.Heptane3.360
12.Toluene4.131
image
GC_FS0563
ColumnRxi®-624Sil MS, 30 m, 0.25 mm ID, 1.40 µm (cat.# 13868)
SampleResidual solvent mix
Diluent:Dimethyl sulfoxide (DMSO)
Conc.:25 ppm (For the HS-FET technique, 10 µL of a 50 µg/mL standard was placed into a 20 mL headspace vial to represent a 25 ppm sample concentration, assuming a 20 mg sample weight.)
Injectionheadspace-loop split (split ratio 10:1)
Liner:Premium 1.0 mm ID straight inlet liner (cat.# 23333.1)
Headspace-Loop
Inj. Port Temp.:250 °C
Instrument:Tekmar HT3
Inj. Time:1.0 min
Transfer Line Temp.:160 °C
Valve Oven Temp.:160 °C
Needle Temp.:140 °C
Sample Temp.:140 °C
Platen temp equil. time:1.0 min
Sample Equil. Time:30.0 min
Vial Pressure:20 psi
Pressurize Time:5.0 min
Loop Pressure:15 psi
Loop Fill Time:2.0 min
Oven
Oven Temp.:35 °C (hold 1.5 min) to 300 °C at 30 °C/min (hold 2.0 min)
Carrier GasHe, constant flow
Linear Velocity:80 cm/sec
DetectorFID @ 320 °C
Make-up Gas Flow Rate:45 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesThe butane used for standard preparation was a mixture of butane and isobutane in an unknown ratio. The concentrations should be considered approximate, but do not exceed 50 ppm for any component.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

residual solvents

,

process solvents

,

Marjuanna,

,

624Sil,

,

Cannabis

,

concentrates

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.