Cannabinoids in Cannabis Sample 94 on Rxi-35Sil MS

PeakstR (sec)Conc. (%)
1.Cannabichromene82.0-
2.Cannabidiol84.65.7
3.delta-8-Tetrahydrocannabinol93.5-
4.delta-9-Tetrahydrocannabinol97.12.0
5.Cannabigerol99.4-
6.Cannabinol104.9-
image
GC_FS0527
ColumnRxi®-35Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13820)
SampleCannabis sample 94
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision® liner w/wool (cat.# 23305.5)
Inj. Temp.:250 °C
Split Vent Flow Rate:25 mL/min
Oven
Oven Temp.:225 °C (hold 0.1 min) to 330 °C at 35 °C/min (hold 0.9 min)
Carrier GasH2, constant flow
Flow Rate:2.5 mL/min
DetectorFID @ 350 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
AcknowledgementThe Ferguson Township Police Department supplied seized marijuana and oversaw sample handling. Frank Dorman at The Pennsylvania State University assisted with sample extraction.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Medical cannabis

,

Cannabidiol

,

Cannabichromene

,

Cannabigerol

,

delta-9-THC

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.