MOSH/MOAH Compounds on Rxi-5Sil MS and Hydroguard-Treated MXT With Allure Silica (Online LC/GC Coupling)

image
GC_FS0500
ColumnRxi-5Sil MS, 15 m, 0.25 mm ID, 0.25 µm (cat.# 13620)
using Hydroguard-Treated MXT Guard Column & Transfer Line 10 m, 0.53 mm ID (cat.# 70084)
Sample10 g rice extracted with 10 mL hexane, evaporated to a final volume of 0.5 mL
InjectionDirect
Oven
Oven Temp.:60 °C (hold 8 min) to 85 °C at 5 °C/min (hold 2 min) to 325 °C at 25 °C/min (hold 6.4 min)
Carrier GasHe, constant pressure (29.01 psi, 200.0 kPa)
DetectorFID
InstrumentDANI Master GC
NotesSample Preparation:
The sample for this chromatogram was fractionated using HPLC under the following conditions:
Column: Allure Silica (cat.# 9160572); Dimensions: 250 mm x 2.1 mm ID; Particle Size: 5 µm; Pore Size: 60 Å; Mobile Phase: A: Hexane, B: Dichloromethane; Gradient (%B): 0.0 min (0% @ 0.3 mL/min), 1.5 min (35% @ 0.3 mL/min), 6.2 min (35% @ 0.3 mL/min), 6.3 min (100% @ 0.5 mL/min), 15.0 min (100% @ 0.5 mL/min), 15.1 min (0% @ 0.5 mL/min), 25.0 min (0% @ 0.5 mL/min), 25.1 min (0% @ 0.3 mL/min), 30.0 min (0% @ 0.3 mL/min); Instrument: Agilent 1260 Infinity.
AcknowledgementRestek thanks Axel Semrau GmbH & Co. KG with their collaborator, DANI Instruments S.p.A., for this chromatogram.
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.