Benz[a]anthracene and Chrysene in a QuEChERS Extract of Mate Tea Separated from Potential Interfering PAHs on Rxi-PAH

PeakstR (sec)
1.Benz[a]anthracene1,028.4
2.Cyclopenta[cd]pyrene1,044.0
3.Triphenylene1,050.0
4.Chrysene1,054.8
image
GC_FF1244
ColumnRxi®-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
Injection
Inj. Vol.:2.5 µL splitless (hold 1 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.5)
Inj. Temp.:275 °C
Purge Flow:40 mL/min
Oven
Oven Temp.:80 °C (hold 1 min) to 210 °C at 40 °C/min to 260 °C at 3 °C/min to 350 °C at 11.5 °C/min (hold 6.25 min)
Carrier GasH2, constant flow
Flow Rate:2.4 mL/min
DetectorTOFMS
Transfer Line Temp.:320 °C
Analyzer Type:TOF
Source Temp.:300 °C
Electron Energy:70 eV
Mass Defect:0 mu/100 u
Solvent Delay Time:3.67 min
Tune Type:PFTBA
Ionization Mode:EI
Acquisition Range:45-550 amu
Spectral Acquisition Rate:5 spectra/sec
InstrumentLECO Pegasus® 4D GCxGC-TOFMS
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

PAHs

,

EFSA PAH4

,

Chrysene

,

Triphenylene

,

Benz[a]anthracene

,

Cyclopenta[cd]pyrene

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.