Food Packaging Volatiles by Purge & Trap GCMS on Rtx-5MS, Overwrap, No Seam

Peaks
1.Tetrahydrofuran
2.Ethylbenzene
3.4-Heptanone
4.Styrene
5.Benzaldehyde
image
GC_FF00460
ColumnRtx-5MS, 30 m, 0.25 mm ID, 1.00 µm (cat.# 12653)
Injectionpurge and trap split (split ratio 20:1)
Liner:1mm ID liner
Inj. Temp.:250 °C
Purge and Trap
Instrument:Tekmar LSC-3100 Purge and Trap
Trap Type:Vocarb 3000, type K
Purge: 10 min @ 60 °C, flow 40 mL/min
Dry Purge: 3 min, flow 40 mL/min
Desorb Preheat Temp.:220 °C
Desorb: 2 min @ 245 °C, flow 40 mL/min
Bake: 6 min @ 230 °C
Interface Connection:injection port
Oven
Oven Temp.:50 °C to 92 °C at 3 °C/min to 220 °C at 20 °C/min (hold 1 min)
Carrier GasHe, constant flow
Flow Rate:1 mL/min
Linear Velocity:36 cm/sec
DetectorMS
Mode:Scan
Source Temp.:280 °C
Electron Energy:70 eV
Ionization Mode:EI
Scan Range:35-260 amu
InstrumentHP6890 GC & 5973 MSD
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.