EPA Method 525.3 OPP Calibration Standard (with 525.3 Internal and Surrogate Standards) on Rxi-5Sil MS

PeakstR (min)
1.Diisopropyl methylphosphonate (DIMP)5.16
2.1,3-Dimethyl-2-nitrobenzene (SS) 7.03
3.Dichlorvos7.80
4.Mevinphos I10.20
5.Mevinphos II10.25
6.Acenaphthene-D10 (IS) 11.08
7.Ethoprop12.83
8.Phorate13.50
9.Dimethipin14.09
10.Pentachlorophenol-C13 (IS)14.19
11.Phenanthrene-D10 (IS) 14.60
12.Disulfoton14.78
PeakstR (min)
13.Phosphamidon15.37
14.Methyl parathion15.71
15.Chlorpyrifos16.65
16.Ethyl parathion16.81
17.Chlorfenvinphos17.64
18.Tetrachlorvinphos18.27
19.Profenofos18.91
20.Tribufos (+merphos)19.06
21.Ethion20.02
22.Triphenyl phosphate (SS) 21.38
23.Chrysene-D12 (IS) 22.21
24.Benzo[a]pyrene-D12 (SS) 26.27
image
GC_EV1432
ColumnRxi-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
SampleEPA Method 525.3 OPP cal standard (cat.# 32543)
EPA Method 525.3 PAH IS mix (cat.# 32547)
EPA Method 525.3 PCP IS (cat.# 32548)
EPA Method 525.3 surrogate standard (cat.# 32549)
Diluent:Ethyl acetate
Conc.:2 µg/mL
Injection
Inj. Vol.:1 µL pulsed splitless (hold 1 min)
Liner:Premium 4 mm single taper w/wool (cat.# 23303.1)
Inj. Temp.:275 °C
Pulse Pressure:30 psi (206.8kPa)
Pulse Time:1 min
Purge Flow:80 mL/min
Oven
Oven Temp.:70 °C (hold 1 min) to 200 °C at 10 °C/min to 320 °C at 7 °C/min (hold 3 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Scan Program:
GroupStart Time
(min)
Scan Range
(amu)
Scan Rate
(scans/sec)
1545-5505.5
Transfer Line Temp.:280 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:6 mm ID
Source Temp.:350 °C
Quad Temp.:200 °C
Solvent Delay Time:5 min
Tune Type:DFTPP
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
Download PDF

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.