Resolution of Benzo Fluoranthene and Benzo Pyrene Isomers on Rxi-PAH (60 m x 0.25 mm x 0.10 µm)

PeakstR (min)Conc.
(µg/mL)
Quant Ion
1.Benzo[b]fluoranthene63.8610252
2.Benzo[k]fluoranthene64.138.3252
3.Benzo[j]fluoranthene64.296.1252
4.Benzo[a]fluoranthene65.100.99252
5.Benzo[e]pyrene67.152.0252
6.Benzo[a]pyrene67.638.9252
7.Perylene68.621.9252
image
GC_EV1423
ColumnRxi-PAH, 60 m, 0.25 mm ID, 0.10 µm (cat.# 49317)
SampleSV internal standard mix (cat.# 31206)
Coronene-D12 (CIL) (cat.# DLM-2715)
Benzo[a]pyrene-D12 (CIL) (cat.# DLM-258-0)
Aromatics in toluene (NIST) (cat.# 2260a)
Native PAH stock (Wellington Labs) (cat.# PAH-STK-A)
EU 15+1 PAH standard (cat.# 32470)
Custom PAH SIM standard (cat.# 557484)
Diluent:Toluene
Conc.:0.71 to 10 µg/mL
Injection
Inj. Vol.:1 µL split (split ratio 10:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.1)
Inj. Temp.:275 °C
Oven
Oven Temp.:110 °C (hold 1.6 min) to 175 °C at 30 °C/min to 265 °C at 1.6 °C/min to 350 °C at 4 °C/min (hold 15 min)
Carrier GasHe, constant flow
Flow Rate:1.0 mL/min
DetectorMS
Mode:SIM
SIM Program:
GroupStart Time
(min)
Ion(s) (m/z)Dwell (ms)
15.09102.1, 108.1, 128.1, 136.2 20
26.68115.1, 142.1 20
37.5576.1, 141.1, 154.1, 156.2 20
48.7775.6, 152.1 20
59.5676.1, 153.1, 162.2, 164.2 20
610.26155.1, 170.2 20
711.1282.4, 165.1 20
814.27139.1, 184.1 20
917.26152.1, 160.2, 178.1, 188.2 20
1020.5894.6, 165.1, 190.1, 192.1 20
1126.32101.1, 202.1 20
1234.9792.1, 184.1 20
1337.36108.0, 216.0 20
1441.78196.1, 212.2 20
1545.86113.1, 226.1, 228.1 20
1647.86114.0, 228.1 20
1749.49113.1, 120.1, 226.1, 228.1, 240.1 20
1851.24154.1, 252.1 20
1954.16119.8, 242.2 20
2060.13125.1, 126.1, 252.1 20
2164.77126.1, 252.1 20
2266.15125.1, 126.1, 132.1, 252.1, 264.1 20
2368.07125.0, 132.2, 252.1, 264.1 20
2469.10252.1, 268.1 20
2571.92139.1, 139.5, 278.1, 279.1 20
2674.87138.1, 139.1, 276.1, 278.1 20
2775.81138.1, 139.1, 278.1 20
2876.82138.1, 276.1 20
2977.53132.6, 138.1, 267.1, 276.1 20
3080.20151.0, 302.1 20
3183.40150.0, 151.0, 156.1, 300.1, 302.1, 312.1 20
3284.88151.0, 302.1 20
Transfer Line Temp.:320 °C
Analyzer Type:Quadrupole
Source Type:Extractor
Extractor Lens:9 mm ID
Source Temp.:350 °C
Quad Temp.:200 °C
Solvent Delay Time:3 min
Ionization Mode:EI
InstrumentAgilent 7890B GC & 5977A MSD
NotesConditions optimized using EZGC software produce good separation of dibenzo[a,c]anthracene and dibenzo[a,h]anthracene from indeno[1,2,3-cd]pyrene, triphenylene from chrysene, as well as the benzofluoranthene isomers.
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

PAH

,

Indenopyrene

,

dibenzanthracene

,

benzofluoranthenes

,

EFSA 15+1

,

16 priority pollutant PAHs

,

EFSA 4

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.