DRO/ORO Calibration Standard on Rxi-1HT (15 m x 0.25 mm x 0.10 µm)

PeakstR (sec)
1.C10118.7
2.C12195.2
3.C14269.6
4.C16337.6
5.C18399.2
image
GC_EV1366
ColumnRxi®-1HT, 15 m, 0.25 mm ID, 0.10 µm (cat.# 13950)
SampleDRO/ORO calibration standard (cat.# 31831)
Diluent:Methylene chloride
Conc.:25,000 ng/µL
Injection
Inj. Vol.:1 µL split (split ratio 50:1)
Liner:Premium 4 mm Precision liner w/wool (cat.# 23305.5)
Inj. Temp.:275 °C
Split Vent Flow Rate:87.5 mL/min
Oven
Oven Temp.:40 °C (hold 0.1 min) to 400 °C at 20 °C/min (hold 1.9 min)
Carrier GasH2, constant flow
Flow Rate:1.75 mL/min
DetectorFID @ 420 °C
Constant Column + Constant Make-up:50 mL/min
Make-up Gas Type:N2
Hydrogen flow:40 mL/min
Air flow:450 mL/min
Data Rate:20 Hz
InstrumentAgilent/HP6890 GC
NotesActual column length was calculated to be 15.7 m using holdup time and EZGC® flow calculator (http://www.restek.com/ezgc-mtfc).
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

Total Petroleum Hydrocarbons

,

Total Recoverable Petroleum Hydrocarbons

,

Alkanes

,

Diesel

,

Hump-o-gram

,

Motor Oil

,

Used Motor Oil

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.