Polycyclic Aromatic Hydrocarbons on Rxi-17 column

PeakstR (min)
1.Naphthalene4.70
2.1-Methylnaphthalene5.28
3.2-Methylnapthalene5.46
4.Acenaphthylene6.45
5.Acenaphthene6.60
6.Fluorene7.18
7.Phenanthrene9.10
8.Anthracene9.14
9.Fluoranthene12.50
10.Pyrene13.33
11.Benz[a]anthracene16.32
12.Chrysene16.58
13.Benzo[b]fluoranthene19.70
PeakstR (min)
14.Benzo[k]fluoranthene19.78
15.Benzo(j)fluoranthene19.95
16.Benzo[a]pyrene21.17
17.3-Methylcholanthrene21.97
18.Dibenz(a,h)acridine24.33
19.Dibenz[a,j]acridine24.39
20.Indeno(1,2,3-cd)pyrene25.04
21.Dibenz[a,h]anthracene25.07
22.Benzo[ghi]perylene26.43
23.7H-Dibenzo(c,g)carbazole27.75
24.Dibenzo(a,E)pyrene34.46
25.Dibenzo(a,i)pyrene35.80
26.Dibenzo(a,h)pyrene36.73
image
GC_EV00925
ColumnRxi®-17, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13523)
SampleSV Calibration Mix #5 / 610 PAH Mix (cat.# 31011)
PAH Supplement Mix for Method 8100 (cat.# 31857)
1-methylnaphthalene (cat.# 31283)
2-methylnaphthalene (cat.# 31285)
Conc.:20 µg/mL each component
Injection
Inj. Vol.:1.0 µL pulsed splitless (hold 0.2 min)
Liner:Drilled Uniliner® (hole near top) (cat.# 21055)
Inj. Temp.:300 °C
Pulse Pressure:20 psi (137.9kPa)
Pulse Time:0.3 min
Purge Flow:40 mL/min
Oven
Oven Temp.:90 °C (hold 1.0 min) to 215 °C at 25 °C/min (hold 0.5 min) to 235 °C at 4 °C/min (hold 0 min) to 280 °C at 15 °C/min (hold 0 min) to 320 °C at 4 °C/min (hold 20 min)
Carrier GasHe, constant flow
Flow Rate:1.2 mL/min
DetectorMS
Mode:Scan
Analyzer Type:Quadrupole
Solvent Delay Time:4.0 min
Tune Type:DFTPP
Ionization Mode:EI
Scan Range:50-550 amu
InstrumentHP6890 GC & 5973 MSD
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

PAHs

,

polyaromatic hydrocarbons

,

polycyclic aromatic hydrocarbons

,

polynuclear aromatic hydrocarbons

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.