Extracted Opiates from Blood on Rxi-5Sil MS

Peaks
1.Hydrocodone
2.Codeine
3.Oxycodone
4.Hydromorphone
5.6-Monoacetylmorphine
6.Morphine
* Matrix
image
GC_CF1150
ColumnRxi®-5Sil MS, 30 m, 0.25 mm ID, 0.25 µm (cat.# 13623)
Sample
Diluent:Ethyl acetate
Conc.:100 ng/mL propionic anhydride derivatives
Injection
Inj. Vol.:1 µL splitless (hold 1 min)
Liner:4 mm splitless taper w/wool (cat.# 22405)
Inj. Temp.:250 °C
Purge Flow:100 mL/min
Oven
Oven Temp.:65 °C (hold 1 min) to 315 °C at 15 °C/min
Carrier GasHe, constant flow
Flow Rate:1 mL/min
Linear Velocity:35 cm/sec @ 65 °C
DetectorMS
Mode:SIM
Transfer Line Temp.:250 °C
Analyzer Type:Quadrupole
Solvent Delay Time:7 min
Tune Type:PFTBA
Ionization Mode:EI
NotesOpiates were spiked into a blood sample and extracted by SPE, then derivatized with propionic anhydride.