Sulfur Compounds on Rxi-1ms

Peaks
1.Hydrogen sulfide
2.Carbonyl sulfide
3.Methyl mercaptan
4.Ethyl mercaptan
5.Dimethyl sulfide
image
GC_AR00861
ColumnRxi®-1ms, 30 m, 0.32 mm ID, 4.00 µm (cat.# 13396)
SampleHydrogen sulfide, carbonyl sulfide, methyl mercaptan, ethyl mercaptan, dimethyl sulfide
Diluent:Helium
Conc.:100 ppbv
Injectionsplitless
Inj. Temp.:30 °C
Oven
Oven Temp.:30 °C
Carrier GasHe, constant pressure
Linear Velocity:48 cm/sec @ 30 °C
DetectorSCD @ 800 °C
NotesInjection: sample loop (30°C), 1 mL splitless direct

Sample Storage and Tansfer:
SilcoCan® air monitoring canister with Siltek®-treated 1/4" valve (cat.# 24182-650); Sulfinert®-treated gas sample loop, 1 cc (cat.# 22848); Sulfinert®-treated gas sample loop, 10 cc (custom order)
Download PDF

relativo alle ricerche nella libreria dei cromatogrammi

sulfur

Consigli per la stampa! Per stampare cromatogrammi su tutta la pagina, cambiare le impostazioni di stampa nella finestra di stampa del vostro software PDF reader: 1)"adatta" o "adatta alla pagina", 2)"panorama" per cromatogrammi grandi o "ritratto" per cromatogrammi alti.